![]() Fluorescence Polarization Measures Changes In Molecular Complex |
|||
|
by Xavier Amouretti Both high throughput screening (HTS) and life science research laboratories require methods to analyze an ever-increasing amount of samples. HTS laboratories focus on testing large numbers of compounds for their ability to elicit or prevent a defined response, while life science research requires easy-to-use tools to study molecular interactions.
How FP worksFP is based on the observation that fluorescent molecules in solution, when excited by polarized light, will then emit polarized light. The resultant light is emitted at a different plane than the original excitatory light (Figure 1) due to the rotation of the molecule. All molecules naturally rotate in solution, although at different rates. Rotation may be influenced by solution viscosity, absolute temperature, molecular volume and the gas constant. If viscosity and temperature are constant through the sample volume, then molecular volume or molecular weight becomes the key variable associated with rotation speed. Large molecules tend to rotate slowly while small molecules rotate rapidly, as the rate of rotation is inversely proportional to a molecule's size.In a typical FP assay, a suitable fluorophore (fluorescent molecule) is bound to a molecule, forming a tracer molecule. As the tracer interacts with other molecules, it may bind to form larger complexes that rotate at a slower rate. This change in rotation speed is identified when the fluorophore is excited with a beam of polarized light. If a molecule remains in the same plane during its excited state, emitted light from the fluorophore will remain polarized. If a molecule rotates out of the plane during the excited state, the emitted light is depolarized.
All the molecules in solution are not initially aligned. Those that are aligned perpendicular to the polarized light will not absorb light and therefore will not emit light, and those at other orientations will absorb light to varying degrees depending on their orientation. The end result will be emitted light that has a fluorescence of some degree in the perpendicular plane to the excitation polarization filter, and a limiting polarization of 500 mP. Therefore, polarization values, corrected for background fluorescence, that are greater than 500 mP indicate either the presence of scattered light, or that the instrument is incorrectly calibrated. Why use fluorescence polarization?As fluorescence polarization provides direct and immediate measure of the binding of a tagged molecule to a target molecule, it is ideally suited to the investigation of molecular interactions. FP experiments are performed without solid support, allowing true steady state analysis to be performed without removal of material or extra washing steps. As FP measurements do not adulterate samples, they can be reanalyzed after modifications and additions, providing an obvious time and cost-savings. Reactions are very rapid, and the reagents are stable, so the process is highly conducive to automation methods. FP reactions are therefore often faster, and more cost efficient than heterogeneous methods.FP assays are easy to construct and are unaffected by solution turbidity, addition of dyes, or sample concentration. Only one tracer is required, and crude receptor preparations may be utilized in the process. Finally, FP results are resistant to variations from detection instruments. When to use fluorescence polarization
Additional homogeneous assay technologies exist that also measure changes in the size of a molecular complex such as Fluorescence Resonance Energy Transfer (FRET), but involve a more difficult dual-labeling process. SummaryFluorescence polarization offers a number of benefits to researchers in both life sciences and high throughput screening. It is one of the most simple techniques available to study biomolecular interactions. It provides cost and time savings, without sacrificing quality and allows for rapid, reproducible and precise results, regardless of the type or quantity of sample being tested.About the author(s)Xavier Amouretti is a Product Manager with BioTek Instruments, Inc. More information about fluorescence polarization is available from BioTek Instruments. The company's Synergy 2 Multi-Detection Microplate Reader (Figure 3) instrument is commonly used to measure FP assays. The instrument incorporates Gen5 software, which automatically calculates polarization and corrects for background interference.www.biotek.com |
|||
© 2008 Advantage Business Media All rights reserved. Use of this website is subject to its terms of use. Privacy Policy |