High Throughput “Direct” Serial Dilution

Featured In: Pharma | Robotics

Tuesday, June 26, 2007

Loading...

by Phil Farrelly and P. Robert Reed

Dilution pitfalls
Recent discussions of problems associated with serial dilution have highlighted the need to improve throughput and accuracy so that increased demands for better and more complete IC50 dose-response analyses and other measures of activity can be met. What if a solution could be found that would produce accurate, verifiable dilutions across an entire plate in less than 1 minute per target plate, and which could run multiple target and source sample plates in a single setup? This report describes one possible answer to this need.

Figure 1. The nano25 provides CVs of 10% at 25 nL to <2% at 1 mL.
Serial dilutions are a time-honored means of providing response data over a regular logarithmic series of sample concentrations to aid in predicting the response of reactions to an agent in varying settings. Because desired dose concentrations of the agent compound are usually very small, the relative inaccuracy or inconsistency of traditionally available pipetting tools at such small volumes has required that the small concentrations be achieved by a series of sequential dilutions. This way, the capabilities of the pipetting tool is better fit to the size of the actual aliquot being handled. If done correctly, even nanoliter sample quantities can be correctly dispensed by ordinary pipettors when mixed in a large enough aliquot of solution.

However, the word "correctly" has serious implications and pitfalls. The normal process, using a microplate, is to start by adding a volume of sample large enough to be adequately handled by the pipettor being used, to a volume of diluent in a well. Then, the two are mixed (typically by repeated aspiration/dispense actions), followed by aspirating part of the mixture and adding it to an adjacent well. The process is then repeated for this and subsequent wells, ultimately resulting in a logarithmic series of increasingly dilute sample concentrations across a row of wells, with very small quantity of sample in the final well. Correctly accomplishing this is not easy. There are a number of potential problems which often result in incomplete or inaccurate sample distribution. The first concern is simply adequate mixing of each sample solution before aspirating it for the next well. This can vary by the solubility, viscosity, and temperature of the sample and its relative concentration in that well. Any incompleteness here will be compounded as each subsequent well is processed. There are also often components which incompletely disssolve in the diluent and may not mix at all, further skewing results of the analysis. If a robotic pipettor is being used, the user must assume that the volumes being aspirated/dispensed are within desired bounds, with no assurance that a clogged tip or air leak in the system has not caused a sample-handling error. Finally, the process usually takes a great deal of time, up to 20 minutes for a 96-well plate.

Dispense directly into wells
A recent trial by a major pharmaceutical research lab of a new product from Hudson Control Group (Springfield, New Jersey), the nano25, has shown that these potential problems can be solved by direct addition of prescribed, decreasing amounts of source sample into pre-dispensed volumes of diluent in each well. Direct addition of the sample itself into the series of wells eliminates the need for the mixing necessary to traditional serial dilutions. Sample components that may be insoluble are not lost in the selected diluent. This way the entire desired amount of sample is placed into each well. Sample-handling errors are eliminated by a flow feedback system that keeps a log of the actual amount of sample that was dispensed. Even if an incorrect amount is dispensed, the dose-response curve would still be accurate because that amount would be known and would produce a valid activity measurement. However, because of this feedback loop, variations in volume caused by viscosity or temperature differences among different samples are overcome automatically by the system, producing consistent results regardless of these factors with no special actions required from the user. Finally, the time required to complete a 96-well plate is reduced to approximately 15 seconds dispensing time. Taking account of the time for washing its tips and aspirating the sample from its source plate, the system can produce a fully-diluted 96-well plate in under 1 minute.

Figure 2. Typical results for a serial dilution are illustrated below, where the correlation factor is greater than 0.9996. The sample dispensed was in 100% DMSO. The graph shows Fluorescence units as a function of requested volume. Click to enlarge.
The system being tested comes equipped with an automated stacker that can feed target and source plates directly onto the deck of the nano25. By feeding target plates while the pipettor is washing its tips and aspirating sample, a real throughput of one target plate per minute has been achieved, as the plate feeding and removal time is very nearly equal to the wash/aspirate time, resulting in little net impact from plate movements. Scanning each plate's barcode, then tying that into the nano25's dispense data log, produces a documented record of the actual sample amounts that were dispensed into each target plate, providing the IC50 analysis with the valid data it needs.

Dispensing accuracy is shown in the linearity of the test results from absorbance readings and weight analyses as shown in Figure 2.


About the author(s)
Phil Farrelly is President of, and P. Robert Reed is a scientist with, Hudson Control Group.

More information about serial dilution and related technologies is available from:
Hudson Control Group
973-376-7400
www.hudsoncontrol.com



Join the Discussion
Rate Article:  Average 0 out of 5
register or log in to comment on this article!

0 Comments

Add Comment

Text Only 2000 character limit

Page 1 of 1

Research Exchange

Publishing Data That Conform to the MIQE Guidelines

Jan 22

Minimum information for publication of Quantitative Real-Time PCR Experiments (MIQE) guidelines help researchers design qPCR experiments.

Calibration Free Analysis to Measure the Concentration of Active Proteins

Nov 23 2009

An SPR-based method, Calibration Free Concentration Analysis can be used to accurately determine the concentration of active protein in a sample, relating to the specific binding activity of the protein, and without the need for a standard.

Advances in EMCCD Technology: Making Imaging Less Arbitrary

Advances in EMCCD Technology: Making Imaging Less Arbitrary

Nov 16 2009

Recent advances in EMCCD technology have solved the problem of non-standardized measurement units by using the photoelectron to standardize imaging experiments.

10 Tips for Successful Sample Concentration and Buffer Exchange

10 Tips for Successful Sample Concentration and Buffer Exchange

Nov 6 2009

Centrifugal devices with ultrafiltration (UF) membrane can solve common problems researchers face when working with proteins.

Advantages of Monolithic Laser Combiner Technology in Confocal Microscopy Systems

Jan 6

Fluorescence microscopy techniques require a reliable light source at the desired wavelength or wavelengths, with minimal downtime for maintenance and alignment. Lasers are a popular light source, although the alignment and upkeep of laser combiners is a time-consuming prospect for many users.

Size-Exclusion Chromatography for Purification of Biomolecules

Dec 2 2009

Size-exclusion chromatography (SEC) is a popular method to separate biomolecules based on their size. Primarily, it is applied to the separation of biopolymers such as proteins and nucleic acids, i.e. water-soluble polymers.

Improving Separation During Electrophoresis

Dec 2 2009

SeparateIT gels represent a novel gel matrix for DNA electrophoresis. Gel polymers are arranged in a conceptually different way, in accordance with a new theoretical model of gel electrophoresis.

Improving Quality of ELISA

Dec 2 2009

Using ready-to-use ELISA kits from manufacturers is easy and convenient. Sometimes however, home-made ELISA is required because there is no kit available with the right antibodies or the characteristics of the available kits such as their limits of detection are not appropriate.

Using the Tecan Genesis Workstation to Automate a Cytometric Bead Array (CBA) Immunoassay

Mar 11

The poster describe the process involved in automating a Cytometric Bead Array (CBA) immunoassay developed to measure relative concentrations of serum antibodies against Tetanus (TT), Sperm Whale Myoglobin (SWM) and Keyhole Limpet Hemocyanin (KLH) in KLH-immunized volunteers.

Ensuring Quality in Assays Performed with Automated Liquid Handlers

Feb 2

The focus of this presentation is to highlight the need of ensuring quality in important assays performed with automated liquid handlers. Nearly all assays performed within a laboratory are volume-dependent. In turn, all concentrations of biological and chemical components in these assays, as well as the associated dilution protocols, are volume-dependent. Because analyte concentration is volume-dependent, an assay’s results might be falsely interpreted if liquid handler variability and inaccuracies are unknown or if the system(s) go unchecked for a long period.

Inkjet System for Protein Crystallography

Feb 1

X-ray crystallography is used routinely by scientists to obtain the three dimensional structure of a biological molecule of interest.Such information can be used to determine how a pharmaceutical interacts with a protein target and what changes might improve functionality. However, the crystallization of macromolecules still remains a serious hindrance in structural determination despite impressive advances in screening methods and technologies.

Attention Deficit & Hyperactivity in a Drosophila Memory Mutant

Attention Deficit & Hyperactivity in a Drosophila Memory Mutant

Nov 9 2009

Action selection is modulated by external stimuli either directly or via memory retrieval. In a constantly changing environment, animals have evolved attention-like processes to effectively filter the incoming sensory stream. These attention-like processes, in turn, are modulated by memory. The neurobiological nature of how attention, action selection and memory are inter-connected is unknown. We describe here new phenotypes of the memory mutant radish in the fruit fly Drosophila.

Interparental conflict, parenting, and childhood depression in a diverse urban population: the role of general cognitive style.

14 hours ago

Research on the mechanisms by which interparental conflict (IPC) affects child depression suggests that both parenting and children's conflict appraisals play important roles, but few studies have explored the role of general cognitive style or included both parenting and...

Lagged cells in the inferior colliculus of the awake ferret.

23 hours ago

Abstract Neurons in the primary auditory cortex (AI) encode complex features of the spectral content of sound, such as direction selectivity. Recent findings of temporal symmetry in AI predict a specific organization of the subcortical input into the cortex that contributes to...

Relationship between multiple sources of perceived social support and psychological and academic adjustment in early adolescence: comparisons across gender.

Mar 8

The current study investigated gender differences in the relationship between sources of perceived support (parent, teacher, classmate, friend, school) and psychological and academic adjustment in a sample of 636 (49% male) middle school students. Longitudinal data were...

Involvement of ceramide in ischemic tolerance induced by preconditioning with sublethal oxygen-glucose deprivation in primary cultured cortical neurons of rats.

Feb 25

The complex molecular cascades of ischemic tolerance in brain cells remain unclear. Recently, sphingolipid-related metabolite ceramide has been implicated as a second messenger in many biological functions, including neuronal survival and death. The present study, therefore,...

Prokariotic Cell Collection in Denmark

Nov 6 2009

I would like to know about a prokariotic cell collection in Denmark. Is there a cell bank in this country? I need a Lactobacillus strain for a fermentation assay and this information about the bank is very helpful for me.

Request for Entries

Oct 16 2009

Ask the Experts is your chance to get the answers to questions on applications, materials, methods, processes, and technologies. Email you question to bst_web@advantagemedia.com, and the editors of Bioscience Technology will find an appropriate expert to answer it. Watch this space in the future to see the questions your colleagues are posting.          

STAY INFORMED: SUBSCRIBE TO

Magazine and E-mail Newsletters

Loading...
E-mail:   

MULTIMEDIA

Video:

Neuroscience Diseases of The Brain and How The Mind Emerges

Neuroscience Diseases of The Brain and How The Mind Emerges

Nov 8 2009

Dennis Choi, director of Emory Universitys Neuroscience Center, is renowned for his groundbreaking research on brain and spinal cord injury.

Podcasts:

Allen Institute for Brain Research

Allen Institute for Brain Research

Oct 14 2009

Discussed in this interview are both the mouse brain project and the human cortex project with an emphasis on the importance of these projects to neuroscience research.